Infrared Training Center

Monday, November 13, 2017

Emissivity and Reflected Temperature

What should I enter for my emissivity and reflected temperature values?
First, ask yourself how are you using your thermal imager?  Is it to identify thermal patterns or measure temperatures?  If you are simply trying to locate thermal patterns such as missing insulation or air leakage in a building (what is called qualitative thermography) measuring an exact temperature really isn’t necessary.  In these situations, leave the values at their default settings (typically 0.95 for emissivity and 68F/20C for reflected temperature) and go for it.

If you need to measure an exact temperature of a motor or a bearing (what is called quantitative thermography) then correctly setting emissivity and reflected temperature is a must to get the most accurate reading.  Doing so, however, should only be attempted by properly trained and certified thermographers…those who are the most qualified to measure temperatures with an infrared camera.

Thermography Certification Dates and Locations



What Is the Best Approach Until Then?
We recommend using the following basic guideline for taking simple measurements.  Know there are many other factors you will need to consider which are covered thoroughly in all ITC training classes.  Until then, to get started, let’s first define emissivity and reflected temperature.

What is Emissivity?
Emissivity is how efficiently an object radiates heat.  It’s defined as the ratio of infrared energy emitted by the object, compared to that emitted by an ideal blackbody, if both are at the same temperature.  It is represented as either a percent or a decimal.

Surfaces exhibit emissivity values ranging anywhere from 0.01 to 0.99.  A highly polished metallic surface such as copper or aluminum are often below 0.10 and are practically an infrared mirror.  Heavily oxidized metallic surfaces will have a much higher emissivity (0.6 or greater depending on the surface condition and the amount of oxidation).  Most flat-finish paints are around 0.90 (in long-wave infrared) while human skin and water are about 0.98.

What is Reflected Temperature?
Reflected temperature (also known as background temperature or Treflected) is any thermal radiation originating from other objects that reflects off the target you are measuring.  To properly obtain an accurate surface temperature reading with thermal imaging, this value (along with that of emissivity) must be quantified and programmed into the camera’s “Object Parameters” (or corresponding software if processing a saved image).  This is used so that the software can compensate for, and ignore, the effects of this radiation as it does not relate to the actual surface temperature of the object you are measuring.

For higher emissivity objects, reflected temperature has less influence.  For lower emissivity objects, however, it’s a critical factor that *must be* understood carefully!   As emissivity decreases, what you are measuring (and seeing thermally) is coming more from the surfaces of surrounding objects (including the camera and operator), not the target you are inspecting.

How to Take a Basic Temperature Measurement
The easiest way to get an accurate measurement is to modify the surface with a material that has a high, known, and consistent emissivity value.  Standard electrical tape, with its emissivity of .95, is one such item that works well for this purpose.

If the surface is safe to touch, simply adhere a piece of electrical tape to that object and set the camera’s emissivity value to 0.95.  Next set the reflected temperature to an appropriate value for the environment.  A stable, room temperature environment will provide the best results.  If you’re still unsure what that value might be, we strongly suggest a minimum of Level I Certification training to understand this concept further.

Next, measure the temperature of the tape with the camera’s spot meter or other measurement tool, being sure that the spot meter’s circle (called the reticle) is filled completely by the target.  IMPORTANT SAFETY NOTE: This “tape method” should *only be utilized* on stationary surfaces that can be accessed safely and are okay to touch.  If attempting this type of measurement on an electrical component, the circuit *must* first be shut off and locked-out/tagged-out before proceeding.

What About Measuring Emissivity and Reflected Temperature on Other Surface Types?
For other surface types, or if temperature measurement accuracy on a variety of other objects is important for you and your inspection program, a minimum of a Level I Thermography Certification Training course is required.   Level I will teach you how to correctly adjust emissivity and reflected temperature on a variety of other components including those that are electrically energized or are difficult to access.  Attendees will learn about the proper procedures needed to evaluate both to ensure thermographers are getting the most accurate temperature readings with their thermal imager.

To learn more about these certification classes, as well as upcoming training dates and locations, please visit the Infrared Training Center online at http://www.infraredtraining.com

55 comments:

  1. This might be a silly question but here goes.
    Were in the northwest and the temperature is 30 degrees, no wind (covered area, tarp overhead) and a humidity level of 48.9.
    Were checking a new substation installation, the substation has been at about 90 percent simulated load for three hours. Buss doors have been opened for us to safely take pictures. They were open during the entire warm-up period.
    Copper bussing is covered by a thick insulating material and were interested in all the tie points in the cubicles. As the copper bussing is usually polished will the insulation material affect this? Camera was set for .79 emissivity.
    Now the question: As were standing outside the cubicle in 30 degree weather that’s what I set the camera (T400) for. Interior temperature will of course be much higher due the load on the bussing. So for the most accurate read was I correct?
    Any suggestions?

    ReplyDelete
    Replies
    1. Gary has provide some good points but I think there are some other aspects that should also be considered. I think the questioner was raising the point as to which temperature value to use for setting Trefl, either, that associated with internal ambient reflections, external ambient reflections or perhaps a mixture of the two!
      Well the only way of measuring this would be to the crinkly foil method, but as Gary rightly points out this should never be attempted when there are exposed live surfaces, and if you were to kill the supply you would need to perform the crinkly measurement quickly before internal reflected temperatures diminished too much.

      My suggestion would be to kill supplies and apply high emissivity paint to all busbar connections just enough to take a reliable measurement from.
      The benefits of this approach is as follows:

      1) For high emissivity surfaces (>0.9), Trefl settings have a negligible effect on surface temperature measurement accuracy, particularly as I suspect you can tolerate up to a 10% discrepancy in accuracy for this application.
      2) Also the high emissivity paint would serve to provide reference points for your measurements for now and future inspections.

      One other point to consider is that in addition to determining the operating temperature of the bus system, it is equally important to identify variations in temperature at each joint and comparing same phase joint temperatures. Bus joint temperatures should generally be the same, any differences, may be indicative of a higher resistance joint. The effects of such defects are catastrophic when these systems are subjected to short circuit events.

      the other point to remember is that internal temperatures will be considerably higher once the substation panels have been replaced.

      Hope this helps, and good luck!

      Delete
    2. Hi! I want to determine emissivity of ice so I can't use the above mentioned method "Step 2: Determining the emissivity" because i can't heat it up by 20 degree celsius. so what should i do?

      Delete
    3. An emissivity measurement requires a temperature difference between the target surface and the reflected apparent temperature. In most cases it is easiest to raise the target surface temperature. However, with ice we have the possibility of melting!

      So in this case I would suggest to lower the reflected apparent temperature. Take the ice target outside on a clear crisp night without cloud cover. A clear night has a much lower reflected temperature than the air, so now the ice will be at a higher temperature than the reflected temperature; and the emissivity measurement can be made.

      Delete
    4. Hi Gary! thank you so much for your support.
      now i can create difference between object temperature and ambient temperature by minimizing the reflection.
      is the other procedure same?
      are the external window temperature and atmospheric temperature same?

      Delete
    5. Hi Umer,

      The external optics parameters are to be used only if you are viewing your target through a window of some sort. For your application, I doubt that is the case, so set the transmission at 100% (or 1.0). Leave the window temperature at the default parameters.

      The atmospheric temperature accesses a model to correct for IR transmission loss between the target surface and the infrared camera. The inputs are distance, atmospheric temperature, and relative humidity. To measure the emissivity, I would set the distance to 0 and leave the other parameters at their default values.

      Delete
    6. Hi Gary! Thank you for your continuous support and sharing free webinar with me.
      1. I tried to create artificial night by surrounding the camera and object with a thick black cloth. But ambient temperature is still near the room temperature and it is much more than actual temperature of ice.

      Delete
    7. The reason for doing this on a clear night is
      1. Eliminate solar reflection and solar gain
      2. Provide a very cold background (T reflection) at a temperature MUCH COLDER than your target.

      A black cloth does not provide the COLD background temperature, and will not work.

      You should really get formal training in IR measurement techniques and science, as these measurements are all about the physics of radiation.

      Delete
    8. Hi Gary! I agree with you! Can you suggest me paid or free online course or training that can help me!
      I watched the webinar about emissivity measurement in Plane English that you recommended me.

      Best Regards

      Delete
    9. I would suggest these two courses as starters:
      1. WEB-TH10 Introduction to Level I;
      http://irtraining.inquisiqr4.com/catalog/course.asp?id=8817&cid=1528

      2. WEB-TH40 - Reflected Apparent Temperature Made Simple;
      http://irtraining.inquisiqr4.com/catalog/course.asp?id=14040&cid=1528

      Delete
  2. 1. As the copper bussing is usually polished will the insulation material affect this?

    Big time! Always use the emissivity of the surface you are viewing and measuring; insulation emissivity for the insulation, copper emissivity for the copper.

    2. As were standing outside the cubicle in 30 degree weather that’s what I set the camera (T400) for. Interior temperature will of course be much higher due the load on the bussing. So for the most accurate read was I correct?

    Use aluminum foil in the same plane and orientation of your target in order to measure reflected temperature. Using ambient air temperature as reflected temperature can lead to major errors. BUT REMEMBER SAFETY FIRST. DO NOT USE ALUMINUM FOIL NEAR ENERGIZED ELECTRICAL SYSTEMS.

    ReplyDelete
  3. i am new to all this but could he have put the foil on securely before energizing then take the readings of the foil , remove foil then energize and take another reading, i am curious on which other way you could do this

    ReplyDelete
    Replies
    1. NO, ABSOLUTELY NOT! SAFETY PROTOCOLS PROHIBIT THIS. THAT IS A GOOD WAY TO GET YOURSELF KILLED. IF YOU DON'T UNDERSTAND ELECTRICAL DISTRIBUTION SYSTEMS, YOU SHOULD NOT BE INSPECTING THEM.

      An easy way to get the reflected temperature is to hold the aluminum at arm's length in front of you in the same orientation as the electrical equipment BUT OBSERVING SAFETY DISTANCE PROTOCOLS.

      These techniques and more are all discussed in our training courses.

      Delete
  4. Sounds like a dangerous practise all-round. With the bussdoors being left open any subsequent arcflash wouldn't care whether it was a real or simulated load. You're also working in the open, a bird flying in the cubicle could cause an arcflash.

    ReplyDelete
  5. And.. when the temperature of the object is above 500ºC, how do you measure emissivity?

    ReplyDelete
    Replies
    1. One way that works well is to drill a hole in the material that is at least 7 times deeper than the diameter. If the object has a homogeneous temperature all the way through, the radiation exiting from this hole will have an effective emissivity extremely close to 1.0. Use this hole as your reference instead of electrical tape.

      Just make sure that your IR camera can optically resolve the hole for measurement purposes.

      Delete
    2. Thx Gary for the answer about emissivity measurement in hot objects.

      Delete
  6. Hi Gary, I'm not very familiar with this method. What is meant by "source of reflections"?

    ReplyDelete
    Replies
    1. When dealing with thermal imaging, we have to understand that surface in our environment is emitting "infrared light" based on the object's temperature and it efficiency factor (emissivity).

      Therefore, when an infrared camera is pointed at an object, what the camera "sees" is infrared energy emitted by the object based on its temperature and emissivity and also the infrared energy reflected by the surroundings off the object (source of reflections).

      We need to measure and enter the apparent temperature of the surroundings temperature, so the camera can properly evaluate its data and compute the temperature of the surface we are actually interested in.

      Delete
    2. Thanks Gary. I'm using the FLIR thermal imaging scanner, but I'm not sure if I'm using it properly to determine the emissivity of some samples. The name "reflection source" suggests that it's a source point, but from your description, it seems to be everything surrounding the sample. So can I know the incident angle between my reflection source and my sample? Why must the reflection source and viewing incident angles be the same? How do I make sure that they are the same?

      For one of my samples, once I measure the tape's temperature in #8 of Step 2, I can't bring my sample's temperature down to match that in the tape by adjusting the emissivity; not even when I reach emissivity = 1.00. Does that mean that my sample has an emissivity higher than that of the tape?

      Two more questions: What's the purpose of taking a thermal image ("freeze the image") in #5 of Step 2? Is the device on the right hand side of Figure 1 the thermal imaging scanner?

      Delete
  7. Hi Gary,

    When dealing with metallic targets (i.e. emissivity < .2) what methods are available for getting accurate temperature readings? Could you still use some sort of blackbody-like reference?

    ReplyDelete
    Replies
    1. When dealing with very low emissivities, the best and easiest approach is to raise the emissivity of the metal via a coating of paint or other high emissivity substance.

      In some cases, drilling a hole into the material that is at least 7 times deeper than the diameter can provide a near unity emissivity as a reference (if the material is uniformly heated all the way through). Be aware of spot size considerations when using this method though.

      Delete
    2. Thanks Gary! Can you go into a little bit more detail about how/why the hole would work? I assume it would simulate a blackbody, in which case could you just use a high emissivity coating in one small (hole sized region)of the low emissivity surface as reference rather than a drilled hole?

      Delete
  8. Hi, I hope you guys are still reading the comments on this topic.
    I would like to determine the emissivity of reptiles, preferrably of live animals of course.
    First of all I assume the emissivity will change once the animal is dead and second of course i do not want to sacrifice an animal for that.
    Does anyone have a suggestion?
    Thanks a lot!
    S.

    ReplyDelete
    Replies
    1. Measuring emissivity of an animal's skin, coats, etc. is always a challenge. Since reptiles are ectotherms, we face an even bigger obstacle. Here is a method I would consider on a live animal.

      1. Place a high emissivity material on the animals skin, e.g. Liquid Paper or similar (research toxicity first of course).
      2. Place the animal in a warm environment (out of direct light).
      3. Move the animal to a uniform cold environment (at least 20 C colder) and immediately capture a thermogram. Measure the reflected temperature of the environment before exposing the animal.
      4. From the thermogram and the emissivity reference (Liquid Paper), you will be able to determine the emissivity of the skin.

      One other thing, I am not sure the emissivity of the skin would change that much if the animal is recently deceased, but this needs to be researched.

      Delete
    2. One other thing, the animal must stabilize in the warmer environment before exposure to the cold environment.

      Delete
    3. Thank you Gary, that sounds great! I will try that. Cheers, S.

      Delete
  9. Another question: if I measure surface temperature with a hand-held infrared thermometer (Lafayette TRP-39; sensitivity: 0.1°C; precision: ± 2%) and simultaneously with a thermo-camera (FLIR
    145 B-200; sensitivity: 0.08°C; precision ± 2%; IR image resolution 200x150 pixels) and I get higher values with the camera the higher the surface AND ambient temperature, which measurement is probably more accurate and where lies the problem? Thanks a lot! S.

    ReplyDelete
    Replies
    1. Let both pieces of equipment stabilize in an operating condition in the environment for 30 minutes before taking critical measurements.
      Make sure both have the same emissivity setting and that the spot being measured are both larger than the minimum spot size that can be measured.

      The results should be similar, but I would trust the imager (not having any experience with that TRP-39).

      Delete
    2. Great, thanks Gary!

      Delete
  10. Hi Gary,

    I would like to know of an alternative method of measuring emissivity.
    I usually capture high temperature vessels with refractory inside and depending what im shooting, the temperatures can be well above 300 - 400 degrees Celsius.
    Alot of the time when I get onto a refinery or petrochemical plant to take infrared scans I am not able to measure the reflected apparent temperatures because I simply cannot reach the object, and therefore sticking insulation (electrical) tape onto it would be a greater challenge. My Thermography Teacher mentioned that I could use a contact thermometer to gauge the difference in temperatures and then adjust my emissivity like that? Is this an accurate way of measuring emissivity? He also mentioned that when measuring high temperature equipment that reflected apparent temperature would not be a major concern. I recently did a scan for sasol and found two major problem areas on their riser line, but the temperatures do not tell the correct story as if the shell was 500 degrees celsius then the paint/coating in that specific area would have peeled/burnt off and it is still fine so I am going to need to go back and re scan with a different emissivity (the correct one this time hopefully) I used an emissivity of 0.84 and It could be between 0.7 to 0.9 in my experience but i need to be 100 % certain before submitting my findings and potentially shutting down a portion of the refinery!!!!
    Please could you advise and Im in dire need of help.

    Kind regards,
    Shaun

    ReplyDelete
    Replies
    1. If you trust your contact thermometer, then that is a good way to calculate the emissivity.

      If your target emissivity is fairly high, and the reflected temperature is low, then accuracy on the reflected temperature is not as critical. However if the surface emissivity is low OR its reflecting something hot, a correct reflected temperature is a must.

      Delete
  11. If we know the physical temperature and apparent temperature then how can we calculate the emissivity?

    ReplyDelete
    Replies
    1. The easiest way is to use your camera or IR software to determine this. Any other way becomes more complicated.

      The basic formula used in cameras is:

      S = e x f(Target Temp) + (1 - e) x f(Reflected Temp)

      Where:
      S = IR radiance received by the camera
      e = target emissivity
      f(T) = the calibration curve of your camera for blackbodies vs temperature

      If you know all the other factors, then solve for e.

      Delete
    2. I didn't understand the term f(T), what does it mean? Kindly educate me about the calibration curve?

      Delete
    3. A calibration curve is generated for an infrared camera by pointing the camera at blackbody sources at various temperatures, and then plotting the camera's signal strength (S) vs temperature. The curve then represents a function of Blackbody temperature f (T).

      Delete
    4. From where I can have the calibration curve, or I should draw it by my self?! if so how I am going to know the signal strength of the camera?

      Delete
    5. I would contact the manufacturer of your infrared camera for that information.

      Delete
  12. Hi Gary,

    Why do we set the emissivity to 1 when measuring the reflected temperature of the aluminum foil?

    ReplyDelete
    Replies
    1. I realized that is a dumb question.

      We are assuming the foil is a black body when setting the camera to an emissivity of 1. However, the foil is essentially a perfect reflector, so it really isn't emitting anything.

      If the foil was perfectly encased in a black box, and the foil is not emitting anything, the camera (with an E=1) would not detect any heat from the foil, assuming the foil is at some high temperature. The foil could possibly be emitting a very little amount of heat, but the camera would barely pick it up. BUT, we know the foil is still at a higher temperature.

      Therefore, we would have to lower the E value of the camera to compensate for this drop in emissivity of the foil to try and calculate an accurate temperature.

      Now, the foil in your instructions is not in a black box. It is getting reflected light from all over the place, the main source being the object you want to measure.

      By setting the camera to E=1, and knowing the foil does not really emit anything, the camera will still see the reflected temperature and will NOT compensate by increasing the temperature, since it is assuming the foil is a perfect black body. The camera is assuming that the reflected heat/light is the emitted heat/light by setting E=1. Therefore, you know that the only temperature you are seeing is the reflected temperature.

      Ta Da!

      Now, this is assuming the camera is calibrated properly using a blackbody calibrator. Is this correct?

      Delete
  13. Apparent temperature measurement explained above is same as ambient temperature setting in the thermal imager?

    ReplyDelete
    Replies
    1. Hi Ravi,

      Some cameras refer to the reflected temperature as ambient or background temperature (see your operator's manual to be sure). In this case, the reflected temperature you measure is entered as the ambient temperature in your camera.

      Delete
  14. Hi, Sorry but looks like I am getting confused on different terms here. What is reflected temp, is it the same temp, that we see in camera? what is apparent temperature? and what is background temp. ?

    I am trying to find an answer of background temp. If I am doing Building thermography outside building, what would be my background temp ? Can i use Air temp with air velocity meter as my background temp. ??

    ReplyDelete
  15. Apparent Temperature: When an infrared measurement device emissivity setting is set to 1.0, the resulting temperature is called an apparent temperature. Apparent temperatures are readings NOT compensated for emissivity, and reflected radiation.

    Reflected temperature is the average apparent temperature of all radiation sources impinging on your target and reflected back to the infrared camera. Reflected radiation contributes no information about the actual temperature of your target, and must be subtracted from the total IR signal coming from your target (performed in your camera or IR software). Background temperature is another term for Reflected temperature.

    Reflected temperature is NOT air temperature. The easiest way to measure this for building analysis is to place a piece of aluminum foil next to a wall or surface of your building, and measure its apparent temperature.

    ReplyDelete
  16. HI
    I HAVE A QUESTION,AM USING FLIR INSTRUMENT I DON'T KNOW HOW TO PUT REFLECTED TEMPERATURE AND DISTANCE BETWEEN THE CAM AND OBJECT.

    ReplyDelete
    Replies
    1. You need to locate the "Parameters" menu on your camera. It is accessed differently on different cameras. You can download a manual for your particular camera at http://flir.custhelp.com/

      Delete
  17. Hi!
    in Infrared Thermography, before performing temperature measurement of different objects we have to enter IRObject Parameters(ambient temperature, distance of object, emissivity etc). but mostly in an IR image we have more than one objects with different emissivities, so the entered emissivity value should relate to which object?

    ReplyDelete
    Replies
    1. I am trying to measure the emissivity of ice by comparing with some standard emissivity object. But I am not able to enter the emissivity of the standard i.e. painted stainless steel thermos and Similarly i can't change the emissivity of ice to ahieve the same temperature.
      can anyone guide me how i can perform this?
      Best Regards
      Umer Sohail

      Delete
    2. The emissivity should be for the particular object surface you want to measure. Some infrared cameras and software allow a separate emissivity to be entered for each measurement tool on the image.

      Delete
    3. To learn more about emissivity and measurement techniques, take this short free online course,
      WEB-TH55 - Emissivity Explained in Plain English
      http://irtraining.inquisiqr4.com/catalog/course.asp?id=51715&cid=1528

      Delete
  18. Hi!
    Can you please tell me the practical calculation of an aluminium object connected to an electrical system??

    ReplyDelete
  19. Hi!

    I wonder about other parameters like distance, temperature, humidity and similar when you determinating reflection temperature? What are the values of them?

    Thank you for your answer.

    Valentina

    ReplyDelete
    Replies
    1. Set distance to 0. This automatically disables the atmospheric transmission model using the air temperature and humidity.

      Delete